Autoclave 180L AT-100

Advantage:

PLC control, touch screen display

Steam-water inner circulation, without steam blowing off

Multiple safety protection devices

Pulse vacuum many times

Introduction:

Fully Automatic Horizontal Pulse Vacuum Autoclave is suitable for sterilizing unwrapped, wrapped, and solid, hollows A , hollow B, instrument with built-in pipe and implanted instrument which can withstand high temperature and pressure in the places including oral department, ophthalmology, operation room and lab.

Technology advantage:

PLC control, color touch screen, the system is powerful and reliable.
Water injection pump to fill water, accurate control, easy to operate and maintenance.
Water filling, pulsing, heating up, sterilizing, drying process are fully automatic without artificial regulation.
Built-in high quality vacuum pulse, vacuum pulse 1~6 times can reach -0.09 Mpa
(-0.9Bar), to ensure effective removal of the air so as to get the optimum saturate steam.
Inner circulation, water ring pump without external pipeline, easy to install.
Unique pulse vacuum drying procedure is designed, and the residue moisture shall be less than 0.2%.
Perfect fault warning system is designed, and fault messages are shown on touch screen display real-timely.
Equipped with mini-printer is to restore sterilization cycle.
Cycle count function.
The beep will sound once the sterilization cycle completes and messages are shown on screen touch display at the same time.

Safety feature

Heater protection: the system will cut off its power supply once the temperature exceeds the desired value and show the alarm information.

Anti-dry heating: anti-dry heating device will automatically cut off its power supply and show the alarm information once the water level is lower than the heating pipe.

Mechanical pressure interlock door: the mechanical door device is locked tightly and can’t be opened once pressure is in chamber.

Over-temperature and over-pressure protection: the system will start up procedures to lower temperature and release pressure to protect items from being damaged when the temperature and pressure in chamber exceed the safe limit.

Automatic release safety valve: the safety valve will begin to release pressure automatically when the pressure in chamber exceeds the desired value.

Electric safety protective measure: short circuit and leakage current are designed to protect equipment and personal safety.

Main parameter:

Designed pressure: 0.26MPa
Designed temperature: 150°C
Max. Working pressure: 0.23MPa
Sterilization temperature range: 105°C~138°C
Temperature accurate: 0.1°C
Temperature homogeneity: ≤±0.5°C
Chamber material: SUS304 stainless steel
Water requirement: distilled water or pure water